Specifications
36
| BRM2 Brain Monitor Reference Manual
EEG specifications
Artifact Rejection:
Automatic
Sensitivity:
±
1 mV for maximum DAU output
Dynamic Range:
0.1
μ
V-1 mV (2-20 Hz)
Update Rate:
0.5 sec (EEG Waveform, Impedance)
1 minute (other parameters)
Event Markers:
User-selected
Impedance Checking:
Automatic
Check signals:
1.35 nArms @ 80, 85, 90 and 95 Hz
aEEG Filter:
0-2 Hz:
2-12 Hz:
12-16 Hz:
16-30 Hz:
60 dB/decade
12 dB/decade
1 dB > 10 Hz level
-120 dB/decade
Spectral Edge:
90% for 2-20 Hz bandwidth
Computed parameters
•
Cross-cerebral and Bilateral amplitude-integrated EEG (aEEG) and Integrated Amplitude
•
EEG intensity
•
Impedance and AC Supply Noise
•
Spectral Edge Index (SEI).
Display modes
•
Real-time (EEG Waveform, aEEG, EEG Power Spectrum)
•
10 minute average (SEI, Integrated Amplitude)
•
Long term trend (SEI, Integrated Amplitude, Intensity, Impedance, AC Supply Noise)
•
Rapid numeric (Impedance).