Brainz BRM3 Brain Monitor Reference Manual |
34
Maintenance
To recalibrate the touchscreen during an assessment or when reviewing an assessment:
1.
Touch
Tools
to display the
Tools
menu, then touch
Calibrate Touchscreen
.
2.
Touch
Yes
to confirm that you want to calibrate the touchscreen. The
Touchscreen Calibration
screen is displayed.
3.
Using a stylus or pencil, touch the exact center of the four target points carefully, when they appear successively in
each corner of the screen.
4.
Touch
Press to complete calibration
. The
touchscreen is automatically recalibrated.
If you miss one of the targets, or “double-tap” the screen, do not touch the
Press to complete calibration
button. The
calibration procedure will restart after 10 seconds.
Upgrading software
The BRM3 Brain Monitor software can be easily upgraded in the field, when required.
Upgrades may be distributed by email or internet download or on a USB memory stick. Simple instructions for
use will be provided with each upgrade.
Follow the instructions supplied with the software upgrade, then retain the upgrade file in case it is needed again.
Servicing
Information on troubleshooting, servicing procedures, and spare part details are included in the
BRM3 Brain
Monitor Technical Manual
. For advice concerning servicing or repair of the BRM3 Brain Monitor, contact an
Authorized Service Center. See
Authorized service centers
on page 41.
Warranty repairs must be performed by an Authorized Service Center.
Do not use
malfunctioning equipment.