Introduction
BRM2 Brain Monitor Reference Manual |
9
Cautions
Read all documentation
Read this Reference Manual, and any other operating instructions or other documentation packed with
the BRM2 Brain Monitor, carefully before operating the instrument in a clinical setting. Keep this
Reference Manual for future use.
Do not autoclave
Do not autoclave any component of the BRM2 Brain Monitor, as serious and irreversible damage may
result.
Ventilation
Do not obstruct the passage of air behind or around the Monitor or block the ventilation inlet holes at its
rear.
Conductive parts
Do not allow the conductive parts of the electrodes and associated connectors, including the Reference
(or neutral) electrode, to contact other conductive parts, including ground, while the BRM2 Brain Monitor
is being used for an assessment.
Disconnection of cables or AC supply
Do not disconnect any of the cables connecting the BRM2 Brain Monitor’s components, or turn off the
AC Supply Switch, or disconnect the AC supply, while the BRM2 Brain Monitor is in use. If any of these
conditions occurs accidentally, reconnect the cables or resume the current assessment. See
Restoration
after AC supply interruption
on page 29.
Cleaning the BRM2 Brain Monitor
Do not clean with concentrated bleach, corrosive chemicals or abrasive cleaning compounds.
Servicing the BRM2 Brain Monitor
Do not open the enclosures of the Monitor, Isolation Unit or Data Acquisition Unit, for any reason. If any
component is faulty, contact your healthcare institution's service personnel or an Authorized Service
Center. See
Authorized service centers
on page 41.
Disconnect when not in use
If the equipment is not to be used for a long time, disconnect it from the AC supply outlet to avoid
damage by transient over-voltage.
Electromagnetic compatibility
The BRM2 Brain Monitor complies with the electromagnetic compatibility requirements of
IEC 60601-1-2. Operation of the device may affect, or be affected by, nearby equipment, due to the
effects of electromagnetic interference. If this happens:
•
Increase the separation between the BRM2 Brain Monitor and the other device.
•
Re-route the device cabling.
•
Plug the devices into AC supply outlets on separate circuit branches.
Refer to the
"BRM2 Technical Manual"
for further compliance information and advice relating to
electromagnetic interference.