Specifications
BRM2 Brain Monitor Reference Manual |
37
Data Acquisition Unit specifications
Low pass filter:
-3 dB at 122 Hz
High pass filter:
-3 dB at 0.5 Hz
Analogue to digital converter:
4 channel successive-approximation type
Sampling rate:
512 samples/sec
Resolution:
12 bits at 512 samples/seconds
Input impedance:
33 M
Ω
in parallel with 4.7 nF
Allowed DC offset:
±
0.75 VDC at input
Noise: <
1
μ
Vp-p
Common mode rejection:
>135 dB
Bandwidth:
0.5 - 122 Hz
Environmental specifications
The following environmental conditions must be maintained to avoid damage to the BRM2 Brain
Monitor.
Operation (all components)
Temperature:
0 to 40 °C (32 to 104 °F)
Relative humidity:
25 to 95% at 40 °C (non-condensing)
Shipping and storage (all components)
Temperature:
-20 to 60 °C (-4 to 140 °F)
Relative humidity:
25 to 95% at 40 °C (non-condensing)