5 The Main Menu
76
IDA
-
3106
Narda
5.2.8
Calibrating the internal compass
Main Menu • Settings • Service Menu
This function is used to calibrate the compass built in to the IDA Basic Unit.
This compass is only used for orientation and aligning the maps displayed
on the instrument. The calibration routine reduces any distortion of the
earth’s magnetic field due to ferrous metals inside the instrument.
Calibration is only needed if there is a significant difference between the
indicated and actual North pole. Calibration must be performed outdoors
and away from any metallic objects.
NOTE:
Newer versions of IDA only have a compass built in to the antenna handle,
not in the Basic Unit itself, so these Basic Units do not require calibration.
All the compass functions are available but the orientation and alignment of
the map displayed on the Basic Unit.
To calibrate the compass:
1.
Press the
Calibrate Int. Compass
softkey.
Instructions for calibration are displayed on the screen.
2.
Follow the instructions for calibration.
Battery Voltage
Actual battery voltage
Battery Capacity
Remaining battery capacity
Buffer Battery Voltage
Internal system battery voltage
Free Memory
Free memory space in megabytes, absolute
Free Memory in Percent
Free memory space in percent, relative
Device Calibration OK
Indicates if a valid device calibration has been made
Operating Hours Counter
Number of hours of operation so far
GPS module detected
Indicates that the internal GPS module is working
properly.
Detected GPS Chipset
Indicates the GPS chipset used
Table 15: Description of
Device Diagnostics
list
Information
Description
Содержание IDA
Страница 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Страница 8: ...Contents VI IDA 3106 Narda...
Страница 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Страница 134: ...6 Common Functions 126 IDA 3106 Narda...
Страница 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Страница 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Страница 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Страница 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Страница 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Страница 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Страница 246: ...15 PC Software 238 IDA 3106 Narda...
Страница 254: ...16 Remote Control 246 IDA 3106 Narda...
Страница 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Страница 272: ...17 Specifications 264 IDA 3106 Narda...
Страница 279: ...Narda IDA 3106 271 19 Glossary The meanings of terms and abbreviations are explained in this chapter...
Страница 284: ...19 Glossary 276 IDA 3106 Narda...
Страница 290: ...Index 282 IDA 3106 Narda...
Страница 291: ......