11 I/Q Analyzer Mode
206
IDA
-
3106
Narda
2.
Enter the desired time. This can be done in the following ways:
–
Keys:
Type in a value using the numerical keys and select the units
using the softkeys.
–
Rotary control knob
: Change the value stepwise using the rotary
control.
–
Rotary control knob + arrow keys
: Use the arrow keys to select the
digit that you want to change. Additional digits can be displayed using
the
key. Change the value of the digit using the rotary control knob.
The value is changed with carry-over.
–
Entry in hh:mm:ss format:
Press the
Mode
softkey. Select a section
using the
keys and change the value using the rotary control
knob.
You cannot enter a Time Resolution value. The time resolution is
determined by the CBW (time resolution = 1/CBW)
3.
Press
OK
to complete entry.
11.3.3 Changing the frequency and time resolution of a
spectrogram
I⁄Q Analyzer • View: Sgram Full
I⁄Q Analyzer • View: Sgram Zoom
I⁄Q Analyzer • View: Persistence
To change the frequency resolution of the spectrogram:
A fast Fourier transformation (FFT) converts a signal from the time domain
to the frequency domain. The number of samples determines whether a
high time resolution (e.g. 256 samples) or a high frequency resolution (e.g.
2,048 samples) is achieved.
A spectrogram view is selected.
1.
Press the
Display
softkey.
2.
Press the
FFT Points
softkey, select an item using the rotary control
knob, and press
OK
to set the selection.
To change the overlapping of the spectrogram:
A higher degree of overlapping in the FFT windowing results in better
amplitude accuracy and higher time resolution. It also results in an
increased computation time for the spectrogram. The overlap can be set in
steps of 50 %, 75 % and 87.5 %.
Содержание IDA
Страница 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Страница 8: ...Contents VI IDA 3106 Narda...
Страница 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Страница 134: ...6 Common Functions 126 IDA 3106 Narda...
Страница 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Страница 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Страница 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Страница 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Страница 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Страница 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Страница 246: ...15 PC Software 238 IDA 3106 Narda...
Страница 254: ...16 Remote Control 246 IDA 3106 Narda...
Страница 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Страница 272: ...17 Specifications 264 IDA 3106 Narda...
Страница 279: ...Narda IDA 3106 271 19 Glossary The meanings of terms and abbreviations are explained in this chapter...
Страница 284: ...19 Glossary 276 IDA 3106 Narda...
Страница 290: ...Index 282 IDA 3106 Narda...
Страница 291: ......