10.3 Selecting measurement parameters
Narda
IDA
-
3106
185
10.3.1 Selecting Time Span, Time Resolution
1.
Press the
Time Span
softkey.
Entry is in seconds by default.
2.
Enter the desired time. This can be done in the following ways:
–
Keys:
Type in a value using the numerical keys and select the units
using the softkeys.
–
Rotary control knob
: Change the value stepwise using the rotary
control.
–
Rotary control knob + arrow keys
: Use the arrow keys to select the
digit that you want to change. Additional digits can be displayed using
the
key. Change the value of the digit using the rotary control knob.
The value is changed with carry-over.
–
Entry in hh:mm:ss format:
Press the
Mode
softkey. Select a section
using the
keys and change the value using the rotary control
knob.
3.
Press the
Time Resolution
softkey and enter the time in the same way.
You can only enter the Time Resolution for the
Long-Time Scope
. In
all other cases, the time resolution is determined by the CBW (time
resolution = 1/CBW)
4.
Press
OK
to complete entry.
10.3.2 Setting the channel bandwidth (CBW)
You must choose a channel bandwidth that allows the entire signal to be
captured. The maximum possible time resolution (1/CBW) results from the
bandwidth setting.
Содержание IDA
Страница 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Страница 8: ...Contents VI IDA 3106 Narda...
Страница 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Страница 134: ...6 Common Functions 126 IDA 3106 Narda...
Страница 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Страница 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Страница 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Страница 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Страница 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Страница 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Страница 246: ...15 PC Software 238 IDA 3106 Narda...
Страница 254: ...16 Remote Control 246 IDA 3106 Narda...
Страница 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Страница 272: ...17 Specifications 264 IDA 3106 Narda...
Страница 279: ...Narda IDA 3106 271 19 Glossary The meanings of terms and abbreviations are explained in this chapter...
Страница 284: ...19 Glossary 276 IDA 3106 Narda...
Страница 290: ...Index 282 IDA 3106 Narda...
Страница 291: ......