12 Multi
-
Channel Power Mode
214
IDA
-
3106
Narda
Setting the RBW manually for the entire table
To set the resolution bandwidth manually:
1.
Press the
BW
softkey.
2.
Use the rotary control knob to highlight the desired value.
(The largest value displayed is the only largest possible value of all
channels shown.)
3.
Press the
OK
key.
NOTE:
Finer resolution normally means slower measurement speed. Regardless of
the selected value, the narrowest resolution bandwidth is limited by the
narrowest channel:
Setting the RBW separately for each channel
NOTE:
Separate RBWs cannot be selected unless
Others = OFF
.
The value selected manually on the device applies to the entire table. You
can also enter the resolution bandwidth for each channel separately using
the PC software. Refer to the online Help function of the PC software for
information on how to do this.
Tip:
More information about the RBW is found in the
12.2.5 Measuring in the gaps between defined
channels (Others)
The field strengths of the channels not selected in the multi-channel table
can be measured and their overall value shown by activating the
Others
function.
NOTE:
Only those other channels between the first and last channels to be
measured are taken into account.
To activate and deactivate the function:
Press the
Others
key until the desired status (
On
,
Off
) is displayed.
The table view shows
Others
; the graphical view shows
O
.
Tip:
The measurement can be speeded up by deactivating the
Others
function.
RBW Fspan narrowest channel
4
------------------------------------------------------------------------------------
Содержание IDA
Страница 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Страница 8: ...Contents VI IDA 3106 Narda...
Страница 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Страница 134: ...6 Common Functions 126 IDA 3106 Narda...
Страница 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Страница 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Страница 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Страница 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Страница 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Страница 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Страница 246: ...15 PC Software 238 IDA 3106 Narda...
Страница 254: ...16 Remote Control 246 IDA 3106 Narda...
Страница 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Страница 272: ...17 Specifications 264 IDA 3106 Narda...
Страница 279: ...Narda IDA 3106 271 19 Glossary The meanings of terms and abbreviations are explained in this chapter...
Страница 284: ...19 Glossary 276 IDA 3106 Narda...
Страница 290: ...Index 282 IDA 3106 Narda...
Страница 291: ......