17.2 IDA
-
3106 Basic Unit
Narda
IDA
-
3106
251
Units
• Anytime: dBm, dBV, dBmV, dBµV
• With Antenna, additionally: V/m, A/m, W/m², mW/cm², dBV/m,
dBmV/m, dBA/m, dBµV/m
Display functions
• Y-scale reference: -130 dBm to 40 dBm
• Y-scale range: 20 dB, 40 dB, 60 dB, 80 dB, 100 dB, 120 dB
• Y-scale auto: automatic scaling
Marker functions
For graphical analysis of Spectrum, Spectrogram, Time Domain
(Scope), I/Q Analyzer, MCP Bar Graph
• Single marker or Delta marker
• Peak Marker: Highest, next, left, right. Adjustable peak
threshold and excursion.
• Peak tracking (selectable)
Demodulation
• Modulation Types: AM, FM, LSB, USB, CW
(Spectrum, Level Meter and Direction Finding modes)
Demodulation bandwidth 100 Hz to 200 kHz
(max. 16 kHz for LSB, USB)
• Representation: Instrument speaker or external earphone
• Squelch: -120 dB to -40 dB nominal, Off
• Audio recording: 16 kHz / 16 bit wave file recording (WAV)
Setups
Up to 200 complete device configurations can be saved in the
basic unit; can be uploaded and downloaded using IDA-3106
Tools software.
Storing
Format
Result can be stored as a screenshot (in png format) or as ASCII
data sets for further evaluation and import into e.g. MS-Excel
Event Controlled Storing
Auto Save function for up to 500 triggered events in
Time Domain (Scope) and I⁄Q Analyzer mode
Time Controlled Storing
• Timer controlled storage of results for long term monitoring (only
Spectrum, Level Meter and Multi
-
Channel Power) start date,
start time and duration: resolution one second.
• Duration maximum 99 hours
• Storage rate settable to every 6 s, 12 s, 18 s, 30 s, 1 min, 2 min,
3 min, 5 min, 6 min, 10 min, 15 min, 20 min, 30 min, 60 min
Memory capacity
128 MB internal memory to store up to 8000 spectra,
4000 screen shots
Operating
temperature range
• -10 °C to +55 °C with battery
• 0 °C to +40 °C with external power supply
Compliance
Climatic
• Storage: 1K3 (IEC 60721-3) extended to -10 °C to +50 °C
• Transport: 2K4 (IEC 60721-3) restricted -30°C to + 70°C due to
display
• Operating: 7K2 (IEC 60721-3) extended to -10 °C to +50 °C
Mechanical
• Storage: 1M3 (IEC 60721-3)
• Transport: 2M3 (IEC 60721-3)
• Operating: 7M3 (IEC 60721-3)
Содержание IDA
Страница 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Страница 8: ...Contents VI IDA 3106 Narda...
Страница 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Страница 134: ...6 Common Functions 126 IDA 3106 Narda...
Страница 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Страница 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Страница 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Страница 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Страница 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Страница 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Страница 246: ...15 PC Software 238 IDA 3106 Narda...
Страница 254: ...16 Remote Control 246 IDA 3106 Narda...
Страница 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Страница 272: ...17 Specifications 264 IDA 3106 Narda...
Страница 279: ...Narda IDA 3106 271 19 Glossary The meanings of terms and abbreviations are explained in this chapter...
Страница 284: ...19 Glossary 276 IDA 3106 Narda...
Страница 290: ...Index 282 IDA 3106 Narda...
Страница 291: ......