17.3 Operating modes
Narda
IDA
-
3106
253
17.3.2 Direction Finding
Measurement principle
Selective level measurement at a selectable frequency
from 9 kHz to 6 GHz.
Possible parameters and settings as specified under
Level Meter.
Antenna direction indication
Numerical display of Azimuth, Elevation and Polarization
determined by the embedded electronic compass of the antenna
handle.
Position indication
Outdoor
Instrument position displayed as latitude and longitude
(GPS WGS84) determined by the embedded GPS receiver of the
basic unit.
Optional:
Graphical indication of the current position drawn on a map.
Indoor
Instrument position set manually on an editable rectangular room
layout.
Detector
Peak or RMS detection
RMS averaging time: selectable, 0.12 s to 30 min
Peak hold time: 480 ms
Display modes
Manual Bearing
Bar graph and numerical display of the signal level and indication
of the direction.
Horizontal Scan
Polar diagram of the signal level vs. antenna orientation,
normalized to the maximum signal. Automatic direction finding and
indication.
Map
Graphical indication of the triangulation results for all
measurement positions. Accepts measurements being performed
by Manual Bearing or Horizontal Scan. Display of the estimated
emitter coordinates, optional drawn on a map (Mapping option).
Horizontal Scan
Continuous
Every 120 ms the polar diagram is updated with the current signal
level and compass data. Start and Stop is initiated by key press on
the antenna handle. The duration of a scan is limited to a
maximum of 4 min. The target azimuth is calculated automatically.
Discrete
For every key press on the antenna handle the polar diagram is
updated with the current signal level and compass data. At least
3 samples are required for calculating the target azimuth (up to
2000 samples are possible). Useful for longer averaging times.
Discrete with Max Hold
Same as for Discrete but using the highest measured values for
direction finding, ideal for intermittent signals.
Содержание IDA
Страница 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Страница 8: ...Contents VI IDA 3106 Narda...
Страница 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Страница 134: ...6 Common Functions 126 IDA 3106 Narda...
Страница 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Страница 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Страница 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Страница 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Страница 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Страница 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Страница 246: ...15 PC Software 238 IDA 3106 Narda...
Страница 254: ...16 Remote Control 246 IDA 3106 Narda...
Страница 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Страница 272: ...17 Specifications 264 IDA 3106 Narda...
Страница 279: ...Narda IDA 3106 271 19 Glossary The meanings of terms and abbreviations are explained in this chapter...
Страница 284: ...19 Glossary 276 IDA 3106 Narda...
Страница 290: ...Index 282 IDA 3106 Narda...
Страница 291: ......