System Description
22
Senscient ELDS™
Rev 15
GB
In this case both an ELDS Transmitter and Receiver has been connected, the ‘active’ connection
is to the Transmitter (the TX button is green and larger). To select the Receiver simply use the
mouse or touchpad to select (click on) the RX button which will then change to green and become
larger.
2.7.3 Installation
The installation button allows an ELDS Transmitter (see section 4.5.1 for more detail) or Receiver
(see section 4.5.2 for more detail) to be installed.
It is essential to install an ELDS Transmitter first when installing an ELDS system. This is
because essential calibration information must be copied from the Transmitter to the
Receiver, so SITE must access the appropriate Transmitter before any attempt is made to
install a particular Receiver. Once a Transmitter has been installed by SITE it will
remember the details of that unit and so it is not necessary to re-install the Transmitter if
the Receiver needs to be re-installed at a later date.
Select the appropriate unit to be installed (TX or RX) before selecting the Installation option.
2.7.4 Test/Diagnostics
The Test/Diagnostics button provides access to key information and test functions regarding the
connected ELDS unit(s). These functions are described in detail in section 7.2 with specific
information regarding the Receiver and Transmitter being in sections 7.2.1 and 7.2.2 respectively.
This option will provide information on either the Transmitter or Receiver or both depending
on which ELDS units are connected. If only one unit is connected then the whole of the
screen is filled with information regarding just that unit. If both a Transmitter and Receiver
then the screen will split vertically with the Receiver information on the left side and the
Transmitter on the right side.
2.7.5 Event Log
This button provides access to the internal ‘event’ log of the selected ELDS unit.
The ELDS unit selected for interrogation is determined by the TX or RX buttons at the
bottom of the SITE main screen. The larger (green) button identifies the unit that will be
examined.
The following screen is presented and will normally be populated by the most recent 50 events
recorded in the ELDS unit selected.
Содержание Senscient ELDS 1000 series
Страница 1: ...Operating Manual Senscient ELDS 1000 2000 Series Open Path Cross Duct Gas Detectors A 5170 0 Rev 15 ...
Страница 58: ...Installation Commissioning 58 Senscient ELDS Rev 15 GB 2 x Mounting Brackets ...
Страница 140: ...Certification 140 Senscient ELDS Rev 15 GB 9 1 2 Control Drawing CENELEC ATEX ...
Страница 142: ...Certification 142 Senscient ELDS Rev 15 GB 9 1 4 Control Drawing CSA US Canada ...
Страница 158: ...Appendix D Manufacturer s EC Declarations 158 Senscient ELDS Rev 15 GB 13 Appendix D Manufacturer s EC Declarations ...
Страница 159: ...Appendix D Manufacturer s EC Declarations Senscient ELDS Rev 15 159 GB ...
Страница 160: ...Appendix D Manufacturer s EC Declarations 160 Senscient ELDS Rev 15 GB ...
Страница 184: ...For local MSA contacts please visit us at MSAsafety com Because every life has a purpose ...