USER MANUAL SWG
Page 93 of 129
Position 1: Acid-gas filter (#56795)
Required materials
2x Acid gas filter (#56795), contained in the service- set.
Required tools:
Needle-nose pliers.
Exchange interval:
Exchange necessary, if filter turns from purple to white.
Steps:
1. Remove exhausted acid gas filter from
the clips.
2. Pull the viton tubes from the
exhausted acid gas filter. If it is
necessary, use needle- nose pliers.
3. Plug the viton tubes on the new acid
gas filters. Be sure, that the tubing is
correct. (The single tubes are signed
-
side.)
Steps:
Pull the viton-tubes from the filter unit. If it is necessary, pliers can be used to solve
the tubes from the filter unit.
Remove the exhausted filter unit from the clip.
Plug the viton tubes on the filter unit. Push the filter unit on the clip.
- Socket
+ Socket
Remove
the
exhausted
filter
unit.
Install a new filter
unit.
Содержание SWG 100 SYNGAS
Страница 1: ...SWG 100 SYNGAS USER MANUAL ...
Страница 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Страница 24: ...USER MANUAL SWG Page 24 of 129 ...
Страница 32: ...USER MANUAL SWG ...
Страница 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Страница 75: ...USER MANUAL SWG Page 75 of 129 Leave the menu and safe the adjustments The individual configuration is done ...
Страница 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Страница 99: ...USER MANUAL SWG Page 99 of 129 ...
Страница 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Страница 114: ...USER MANUAL SWG Page 114 of 129 ...
Страница 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Страница 127: ...USER MANUAL SWG Page 127 of 129 ...
Страница 129: ...USER MANUAL SWG Page 129 of 129 ...