USER MANUAL SWG
Page 20 of 129
3.4.
Installation of pressure reducer unit
•
DANGER
Explosion hazard
Do not operate the analyser without pressure
reducer unit.
•
DANGER
Explosion hazard
The measuring point must not have an
overpressure of 150 mbar.
•
WARNING
As part of the safety concept, a reducing unit must be
installed on each Autocal inlet.
•
WARNING
As part of the safety concept, a reducing unit must be
installed on each sample gas inlet.
NOTE
The copper seal of the nozzle is intended only for
single use.
NOTE
Use PTFE strips to tighten the fitting inside the thread!
NOTE
The pressure reducer unit is equipped with a PTFE
DN4/6 fitting. Use a 1/8G outer thread fitting for using
another fitting.
For safety reasons, the gas inlet and the autocalibration inlet must be equipped
with a pressure reducing unit. This fitting prevents the gas flow from being
limited in the event of a pipe leak in the unit. The ventilation of the unit prevents
the creation of an explosive atmosphere. This is only guaranteed if the
overpressure at the measuring point does not exceed 150 mbar.
The sketch below shows how the pressure reducer unit is built.
Содержание SWG 100 SYNGAS
Страница 1: ...SWG 100 SYNGAS USER MANUAL ...
Страница 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Страница 24: ...USER MANUAL SWG Page 24 of 129 ...
Страница 32: ...USER MANUAL SWG ...
Страница 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Страница 75: ...USER MANUAL SWG Page 75 of 129 Leave the menu and safe the adjustments The individual configuration is done ...
Страница 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Страница 99: ...USER MANUAL SWG Page 99 of 129 ...
Страница 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Страница 114: ...USER MANUAL SWG Page 114 of 129 ...
Страница 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Страница 127: ...USER MANUAL SWG Page 127 of 129 ...
Страница 129: ...USER MANUAL SWG Page 129 of 129 ...