USER MANUAL SWG
Page 40 of 129
1.
Example: 4-20 mA signal outputs to control room
Starting position:
A plan wants to read the measurement values from the CO2 channel to their
DCS
1
. The SWG100 has only one IO module and three sample points. All outputs
are free. The CO2 values from the second sample point should be logged in the
control room.
Follow steps must be done to do this:
•
Wiring the analog-output 1 with the DCS: Wiring the analog-
output (AO1+/AO1-) with the DCS. For this, use a wiring with a
conductor cross section between
•
In the next step output 1 must be configured at the SWG100. To
do this, open the menu:
EXTRAS/
CONFIGURATE
ANALOG
OUTPUTS
.
The following menu overview will be appeared (see screenshot
below). To configure the IO1, select the first record (channel I/O
1/1 SP1 / CH4) in the list and open it.
Select
A
NALOG
OUTPUT
and OK
•
The follow screen appears. To configurate the channel 1, the
red marked positions must be changed:
1
DCS=Distributed Control System
DCS
Содержание SWG 100 SYNGAS
Страница 1: ...SWG 100 SYNGAS USER MANUAL ...
Страница 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Страница 24: ...USER MANUAL SWG Page 24 of 129 ...
Страница 32: ...USER MANUAL SWG ...
Страница 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Страница 75: ...USER MANUAL SWG Page 75 of 129 Leave the menu and safe the adjustments The individual configuration is done ...
Страница 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Страница 99: ...USER MANUAL SWG Page 99 of 129 ...
Страница 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Страница 114: ...USER MANUAL SWG Page 114 of 129 ...
Страница 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Страница 127: ...USER MANUAL SWG Page 127 of 129 ...
Страница 129: ...USER MANUAL SWG Page 129 of 129 ...