USER MANUAL SWG
Page 124 of 129
Mounting rules for heated sampling lines
Basic rules for laying the heated sampling lines
Wrong
Right
Do not lay several heated sampling lines
close together. -> Overheating hazard!!
Care for enough space between the
single heated sampling lines.
Do not use clips, being too tight.
→
Crushing hazard!!
Only use clips, which have the right side.
Do not fixing the heated sampling line
with glue or material, which coat the entire
line.->Overheating hazard!!
Lay the heated sampling line free and
clean.
Do not hang the heated sampling line on a
direction change, without a check rail.
Only use direction changes, having the
right check rail diameter.
Do not choose the heated sampling line
too short for your application. -> Could
cause sharp kink!!
The ends of the heated sampling line
need min. 5x the diameter of the
Содержание SWG 100 SYNGAS
Страница 1: ...SWG 100 SYNGAS USER MANUAL ...
Страница 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Страница 24: ...USER MANUAL SWG Page 24 of 129 ...
Страница 32: ...USER MANUAL SWG ...
Страница 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Страница 75: ...USER MANUAL SWG Page 75 of 129 Leave the menu and safe the adjustments The individual configuration is done ...
Страница 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Страница 99: ...USER MANUAL SWG Page 99 of 129 ...
Страница 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Страница 114: ...USER MANUAL SWG Page 114 of 129 ...
Страница 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Страница 127: ...USER MANUAL SWG Page 127 of 129 ...
Страница 129: ...USER MANUAL SWG Page 129 of 129 ...