3.5.
Connection of main power supply
The connection of the main power supply needs a remote power supply box. The shut-
off box must be installed in an area where there is no possibility of an explosive
atmosphere.
Remote Power Supply Box
The shut off box is necessary to power up and down the analyzer in a safe area. It
prevent that a brake away spark is generated in the industrial area. The shut off box
belongs to the safety concept of the analyser.
The Remote Power Supply Box as optionally supplied by MRU, has a switch-off
contactor
(K1)
, with 3 make contacts (normally open). The power contactor K1 is
activated after the analyser is powered on by pushing the power-on button until the
green LED is light-on. In case of an alarm, the power contactor K1 is deenergized. The
circuit diagram shows the correct wiring of the power contactor installed in the
remote power supply box.
The circuit diagram of above pictured device must be implemented by user as part of
the safety concept.
Содержание SWG 100 SYNGAS
Страница 1: ...SWG 100 SYNGAS USER MANUAL ...
Страница 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Страница 24: ...USER MANUAL SWG Page 24 of 129 ...
Страница 32: ...USER MANUAL SWG ...
Страница 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Страница 75: ...USER MANUAL SWG Page 75 of 129 Leave the menu and safe the adjustments The individual configuration is done ...
Страница 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Страница 99: ...USER MANUAL SWG Page 99 of 129 ...
Страница 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Страница 114: ...USER MANUAL SWG Page 114 of 129 ...
Страница 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Страница 127: ...USER MANUAL SWG Page 127 of 129 ...
Страница 129: ...USER MANUAL SWG Page 129 of 129 ...