Example for a measurement cycle configuration
In this chapter an individual measurement cycle should be created with the
features described at the chapters below.
The measurement cycle should have the following sequence:
Following points must be done to configure the individual measurement cycle:
•
Open the measurement cycle config. menu: Path:
E
XTRA
/
MEASUREMENT CYCLE CONFIG
.
•
The default measurement cycle will appear. Open the cycle
phase detail of the first zeroing and adjust the phase duration
at 25 min.
•
Leave the cycle phase detail and select the second point.
Switch with the
left/right arrow keys
until the measurement
SP2 is selected.
•
•
Open with the
OK
key the cycle phase detail of the
measurement SP2. Adjust the duration-phase at 15 min. and
the suction/response time until the pure meas. time has the
value of 3 min. Use for this operation the
arrow keys
.
Содержание SWG 100 SYNGAS
Страница 1: ...SWG 100 SYNGAS USER MANUAL ...
Страница 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Страница 24: ...USER MANUAL SWG Page 24 of 129 ...
Страница 32: ...USER MANUAL SWG ...
Страница 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Страница 75: ...USER MANUAL SWG Page 75 of 129 Leave the menu and safe the adjustments The individual configuration is done ...
Страница 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Страница 99: ...USER MANUAL SWG Page 99 of 129 ...
Страница 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Страница 114: ...USER MANUAL SWG Page 114 of 129 ...
Страница 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Страница 127: ...USER MANUAL SWG Page 127 of 129 ...
Страница 129: ...USER MANUAL SWG Page 129 of 129 ...