10 - 13
10.5.7
Cleaning the TemporalScanner™ Thermometer
WARNING
•
Never use abrasive materials (such as steel wool or silver polish), or erosive
cleaners (such as acetone or acetone-based cleaners).
•
Never immerse any part of the product in liquids or allow liquid to enter the
interior.
•
When cleaning or disinfecting the product, avoid pouring any liquid on the
thermometer.
You should clean the product as per your hospital’s regulations after each use.
Sterilization is not allowed for the product.
■
Cleaning the case
The TemporalScanner™ case can be wiped down using a cloth dampened with
70% isopropyl alcohol. The industrial grade housing and design of the electronic
components allow for completely safe cleaning with 70% isopropyl alcohol but
should not be immersed in fluid or autoclaved.
■
Cleaning the sensor lens
Dirt, greasy films or moisture on the lens will interfere with the passage of infrared
heat and affect the accuracy of the product.
Clean the lens every two weeks with a cotton swab dipped in alcohol. Use only
light force for cleaning, to avoid damaging the lens. Water can be used to remove
any residual film left by the alcohol. Do not use bleach or other cleaning solutions
on the sensor lens.
10.6
Monitoring Temp with HeTaida Thermometer
The HeTaida thermometer is a non-contact infrared thermometer that measures
temporal temperature. It is applicable for adult, pediatric and neonatal patients. The
thermometer connects with the monitor through the connector on the back.
10.6.1
Overview of HeTaida Thermometer
1
2
4
5
3
Содержание VS 8A
Страница 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 8: ...2 6 This page intentionally left blank ...
Страница 20: ...12 D 2 9 Other System Technical Alarm Messages D 10 E Abbreviations E 1 F Declaration of Conformity F 1 ...
Страница 26: ...1 6 This page intentionally left blank ...
Страница 58: ...3 24 This page intentionally left blank ...
Страница 66: ...5 4 This page intentionally left blank ...
Страница 70: ...6 4 This page intentionally left blank ...
Страница 98: ...8 16 This page intentionally left blank ...
Страница 142: ...13 4 This page intentionally left blank ...
Страница 162: ...15 16 This page intentionally left blank ...
Страница 166: ...16 4 This page intentionally left blank ...
Страница 202: ...19 8 This page intentionally left blank ...
Страница 224: ...22 10 This page intentionally left blank ...
Страница 242: ...A 18 A 7 Operating Environment Host CPU AM3358 Primary programming language C Operating system Linux 3 2 0 ...
Страница 250: ...B 8 This page intentionally left blank ...
Страница 260: ...C 10 This page intentionally left blank ...
Страница 271: ...D 11 This page intentionally left blank ...
Страница 272: ...D 12 ...
Страница 275: ...F 1 F Declaration of Conformity ...
Страница 276: ...F 2 This page intentionally left blank ...
Страница 277: ......
Страница 278: ...P N 046 020097 00 4 0 ...