12 - 5
12.7.2
Setting the CO
2
Waveform
In
Continuous Monitoring
mode, you can set the CO
2
waveform. Follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select
Setup
tab.
3.
Set
Waveform Type
,
Speed
and
CO2 Scale
of the CO
2
waveform.
◆
Select
Waveform Type
and toggle between
Draw
and
Fill
.
Draw
: The CO
2
wave is displayed as a curved line.
Fill
: The CO
2
wave is displayed as a filled area.
◆
Select
Speed
and then select the appropriate setting. The faster the wave
sweeps, the wider the wave is.
◆
Select
CO2 Scale
and then change the size of the CO
2
waveform.
12.7.3
Entering the Standby Mode
In
Continuous Monitoring
mode, you can set the CO
2
module to one of the following
modes according to the module status:
■
Select
Measure
mode when you use the CO
2
module for monitoring.
■
Select
Standby
mode when you does not use the CO
2
module to prolong the
service life of the CO
2
module.
The default operating mode is
Measure
. If you are not using the CO
2
module, you can
proceed as follows to enter the Standby mode:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select
Setup
tab.
3.
Set
Operating Mode
to
Standby
.
12.7.4
Setting the Auto Standby
The monitor enters the standby mode automatically after the configured period of time
if no breath is detected since the last detected breath.
To set the auto standby, follow this procedure:
1.
Select the CO
2
parameter area or waveform area to enter the
CO2
menu.
2.
Select
Setup
tab.
3.
Set
Auto Standby
.
12.7.5
Setting Humidity Compensation
The CO
2
modules is configured to compensate CO
2
readings for either Body
Temperature and Pressure, Saturated Gas (BTPS), to account for humidity in the patient’s
breath, or Ambient Temperature and Pressure, Dry Gas (ATPD).
Содержание VS 8A
Страница 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 8: ...2 6 This page intentionally left blank ...
Страница 20: ...12 D 2 9 Other System Technical Alarm Messages D 10 E Abbreviations E 1 F Declaration of Conformity F 1 ...
Страница 26: ...1 6 This page intentionally left blank ...
Страница 58: ...3 24 This page intentionally left blank ...
Страница 66: ...5 4 This page intentionally left blank ...
Страница 70: ...6 4 This page intentionally left blank ...
Страница 98: ...8 16 This page intentionally left blank ...
Страница 142: ...13 4 This page intentionally left blank ...
Страница 162: ...15 16 This page intentionally left blank ...
Страница 166: ...16 4 This page intentionally left blank ...
Страница 202: ...19 8 This page intentionally left blank ...
Страница 224: ...22 10 This page intentionally left blank ...
Страница 242: ...A 18 A 7 Operating Environment Host CPU AM3358 Primary programming language C Operating system Linux 3 2 0 ...
Страница 250: ...B 8 This page intentionally left blank ...
Страница 260: ...C 10 This page intentionally left blank ...
Страница 271: ...D 11 This page intentionally left blank ...
Страница 272: ...D 12 ...
Страница 275: ...F 1 F Declaration of Conformity ...
Страница 276: ...F 2 This page intentionally left blank ...
Страница 277: ......
Страница 278: ...P N 046 020097 00 4 0 ...