8 - 11
3.
Set
Averaging
.
8.6.8
Changing the Sensitivity (for Mindray SpO
2
)
The SpO
2
value displayed on the monitor screen is the average of data collected within a
specific time. The shorter the averaging time is, the quicker the monitor responds to
changes in the patient’s oxygen saturation level. Contrarily, the longer the averaging
time is, the slower the monitor responds to changes in the patient’s oxygen saturation
level, but the SpO
2
measurement is more stable. For critically ill patients, selecting
shorter averaging time will help understanding the patient’s state.
To set the averaging time, follow this procedure:
1.
In
Continuous Monitoring
mode, select the SpO
2
numeric area or waveform area
to enter the
SpO2
menu. In
Spot Check
mode, select the
Main Menu
quick key
→
from the
Parameters
column select
Setup
→
select
SpO2
.
2.
Select the
Setup
tab.
3.
Set the
Sensitivity
.
8.6.9
Showing/Hiding PI
You can set whether to display PI in the SpO
2
parameter area. To do so, follow this
procedure:
1.
In
Continuous Monitoring
mode, select the SpO
2
numeric area or waveform area
to enter the
SpO2
menu. In
Spot Check
mode, select the
Main Menu
quick key
→
from the
Parameters
column select
Setup
→
select
SpO2
.
2.
Select the
Setup
tab.
3.
Switch on or off
Display PI.
NOTE
•
This function is only available for Mindray SpO
2
and Masimo SpO
2
.
8.6.10
Monitoring SpO
2
and NIBP Simultaneously
When monitoring SpO
2
and NIBP on the same limb simultaneously, you can switch on
NIBP Simul
to lock the SpO
2
alarm status until the NIBP measurement ends. If you
switch off
NIBP Simul
, low perfusion caused by NIBP measurement may lead to
inaccurate SpO
2
readings and therefore cause false physiological alarms.
In
Continuous Monitoring
mode, you can set the
NIBP Simul
. Follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
Alarm
tab.
3.
Set
NIBP Simul
.
Содержание VS 8A
Страница 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 8: ...2 6 This page intentionally left blank ...
Страница 20: ...12 D 2 9 Other System Technical Alarm Messages D 10 E Abbreviations E 1 F Declaration of Conformity F 1 ...
Страница 26: ...1 6 This page intentionally left blank ...
Страница 58: ...3 24 This page intentionally left blank ...
Страница 66: ...5 4 This page intentionally left blank ...
Страница 70: ...6 4 This page intentionally left blank ...
Страница 98: ...8 16 This page intentionally left blank ...
Страница 142: ...13 4 This page intentionally left blank ...
Страница 162: ...15 16 This page intentionally left blank ...
Страница 166: ...16 4 This page intentionally left blank ...
Страница 202: ...19 8 This page intentionally left blank ...
Страница 224: ...22 10 This page intentionally left blank ...
Страница 242: ...A 18 A 7 Operating Environment Host CPU AM3358 Primary programming language C Operating system Linux 3 2 0 ...
Страница 250: ...B 8 This page intentionally left blank ...
Страница 260: ...C 10 This page intentionally left blank ...
Страница 271: ...D 11 This page intentionally left blank ...
Страница 272: ...D 12 ...
Страница 275: ...F 1 F Declaration of Conformity ...
Страница 276: ...F 2 This page intentionally left blank ...
Страница 277: ......
Страница 278: ...P N 046 020097 00 4 0 ...