3 - 5
3.4.3
Connecting a Barcode Reader
The monitor supports both linear (1D) barcode reader and two-dimension (2D) barcode
reader. The barcode reader is connected to the monitor via the USB connector.
3.4.4
Connecting a Digital Scale
You can connect a HEALTH O METER 600KL-BT scale via the USB connector to obtain the
height, weight, and BMI data of a patient.
3.5
Turning on the Monitor
Before turning on the monitor, perform the following inspections:
1.
Check the monitor for any mechanical damage. Make sure that all external cables,
plug-ins and accessories are properly connected.
2.
Connect the monitor to the AC power source. Make sure the battery power is
sufficient if the monitor is powered by the battery.
3.
Press the power switch to turn on the monitor.
The monitor automatically performs a self test at startup. Check that the alarm tone is
heard and the alarm lamp illuminates, one after the other, in red, yellow, and cyan. This
indicates that the visible and audible alarm indicators function correctly.
CAUTION
•
Check that visual and auditory alarm signals are presented correctly when
the equipment is powered on. Do not use the patient monitor for any
monitoring procedure on a patient if you suspect it is not working properly,
or if it is mechanically damaged. Contact your service personnel or us.
3.6
Start Wizard
When the monitor is started for the first time, a start wizard is launched to help set up
your monitor. You can:
■
Set the system language: select a language from the list and then select
.
■
Set the Temp type: select the Temp module of your monitor and then select
.
■
Select
Import Configuration
if a preferred configuration has been exported and
saved in a USB drive. For details, refer to
5.5.2 Importing a Configuration
■
Select
Create New Configuration
to:
◆
Set the system time: for details of the settings, see
◆
Set parameter units: for details of the settings, see
Содержание VS 8A
Страница 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 8: ...2 6 This page intentionally left blank ...
Страница 20: ...12 D 2 9 Other System Technical Alarm Messages D 10 E Abbreviations E 1 F Declaration of Conformity F 1 ...
Страница 26: ...1 6 This page intentionally left blank ...
Страница 58: ...3 24 This page intentionally left blank ...
Страница 66: ...5 4 This page intentionally left blank ...
Страница 70: ...6 4 This page intentionally left blank ...
Страница 98: ...8 16 This page intentionally left blank ...
Страница 142: ...13 4 This page intentionally left blank ...
Страница 162: ...15 16 This page intentionally left blank ...
Страница 166: ...16 4 This page intentionally left blank ...
Страница 202: ...19 8 This page intentionally left blank ...
Страница 224: ...22 10 This page intentionally left blank ...
Страница 242: ...A 18 A 7 Operating Environment Host CPU AM3358 Primary programming language C Operating system Linux 3 2 0 ...
Страница 250: ...B 8 This page intentionally left blank ...
Страница 260: ...C 10 This page intentionally left blank ...
Страница 271: ...D 11 This page intentionally left blank ...
Страница 272: ...D 12 ...
Страница 275: ...F 1 F Declaration of Conformity ...
Страница 276: ...F 2 This page intentionally left blank ...
Страница 277: ......
Страница 278: ...P N 046 020097 00 4 0 ...