3 - 20
3.9.9
Checking Software Licenses
To run the following functions in your monitor, software licenses are required:
■
Orthostatic BP measurement
■
Vitalslink
■
Network Connectivity
■
MLDAP
■
SpO2 Calc RR
To check the licenses, select the
Main Menu
quick key
→
select
License
→
Local
.
To install the licenses, follow this procedure:
1.
Connect the USB drive with the licenses in to the monitor’s USB connector.
2.
Select the
Main Menu
quick key
→
select
License
→
select
External
.
3.
Select
Install
.
3.10
Start Working
3.10.1
Selecting a Work Mode
The monitor provides spot check mode and continuous monitoring mode. It works
based on the workflow you selected.
3.10.1.1
Continuous Monitoring Mode
The continuous monitoring mode is used for long-term patient monitoring. To switch to
monitoring mode, select the workflow area and select a workflow under
Continuous
Monitoring
.
3.10.1.2
Spot Check Mode
The spot check mode is intended for on-spot measurement in a short period. To switch
to spot check mode, select the workflow area and select a workflow under
Spot Check
.
In spot check mode, physiological alarm system is disabled.
■
There will be no physiological alarm related indications.
I/O Fluid Unit
ml
,
L
ml
CO2 Unit
mmHg
,
kPa
,
%
mmHg
Temp Unit
°C
,
°F
°C
Pressure Unit
mmHg
,
kPa
mmHg
Parameter
Options
Default
Содержание VS 8A
Страница 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 8: ...2 6 This page intentionally left blank ...
Страница 20: ...12 D 2 9 Other System Technical Alarm Messages D 10 E Abbreviations E 1 F Declaration of Conformity F 1 ...
Страница 26: ...1 6 This page intentionally left blank ...
Страница 58: ...3 24 This page intentionally left blank ...
Страница 66: ...5 4 This page intentionally left blank ...
Страница 70: ...6 4 This page intentionally left blank ...
Страница 98: ...8 16 This page intentionally left blank ...
Страница 142: ...13 4 This page intentionally left blank ...
Страница 162: ...15 16 This page intentionally left blank ...
Страница 166: ...16 4 This page intentionally left blank ...
Страница 202: ...19 8 This page intentionally left blank ...
Страница 224: ...22 10 This page intentionally left blank ...
Страница 242: ...A 18 A 7 Operating Environment Host CPU AM3358 Primary programming language C Operating system Linux 3 2 0 ...
Страница 250: ...B 8 This page intentionally left blank ...
Страница 260: ...C 10 This page intentionally left blank ...
Страница 271: ...D 11 This page intentionally left blank ...
Страница 272: ...D 12 ...
Страница 275: ...F 1 F Declaration of Conformity ...
Страница 276: ...F 2 This page intentionally left blank ...
Страница 277: ......
Страница 278: ...P N 046 020097 00 4 0 ...