D - 4
D.2
Technical Alarm Messages
This section lists technical alarms, their default priority, indication on alarm reset, and the actions that can be
taken when an alarm occurs.
Technical alarms give different alarm indicators when the alarm system is reset. In this section we classify the
technical alarms into three categories for easy clarification:
■
A: technical alarms are cleared. The monitor gives no alarm indications.
■
B: technical alarms are changed to the prompt messages.
■
C: the alarm is silenced and a √ appears before the alarm message.
In the following tables we will use A, B, and C to refer to the indications on alarm reset.
D.2.1
General Technical Alarm Messages
Note: XX represents a measurement or parameter label, such as HR, RR, SpO
2
, EtCO
2
, and so on.
D.2.2
ECG Technical Alarm Messages
Note: XX represents ECG lead name, for example RL, LL, V, Va, Vb, and so on.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
XX Module Error
High
C
XX module does not work properly. Replug the
module, if the alarm persists, contact your service
personnel.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
ECG Noisy
Low/Prompt
A
The ECG signal is noisy. Check for any possible
sources of signal noise around the cable and
electrode, and check the patient for excessive
motion.
ECG Amplitude Too Small
Low
C
The ECG amplitude does not reach the detected
threshold. Check for any possible source of
interference around the cable and electrode.
ECG Lead Off
Low
B
The electrode has become detached from the
patient or the lead wire has become
disconnected from the adapter cable. Check the
connections of the electrodes and leadwires.
ECG XX Lead Off
Low
B
The electrode has become detached from the
patient or the lead wire has become
disconnected from the adapter cable. Check the
connections of the electrodes and leadwires.
ECG Signal Invalid
Low
A
Patient skin impedance is too high. Check ECG
electrode application.
ECG Learning
Prompt
/
ECG learning is manually or automatically
triggered.
Cannot Analyze QT
Prompt
/
/
D12L Not Available
Prompt
C
The current Va and Vb combination does not
support D12L. Choose an available Va and Vb
combination. For more information, see
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...