A - 6
A.11
Data Storage
IBP Analog Output
Bandwidth (-3dB; reference
frequency:1Hz)
0 to 40 Hz
Maximum transmission delay
30 ms
Gain (reference frequency 1 Hz)
1 V/100 mmHg, ±5%
Nurse Call Signal
Amplitude
High level: 3.5 to 5 V, ±5%, providing a minimum of 10 mA output current;
Low level: < 0.5 V, receiving a minimum of 5 mA input current.
Rising and falling time
≤1 ms
Defib Sync Pulse
Output impedance
≤100 ohm
Maximum time delay
35 ms (R-wave peak to leading edge of pulse)
Amplitude
High level: 3.5 to 5 V, ±5%, providing a maximum of 10 mA output current;
Low level: < 0.5 V, receiving a maximum of 5 mA input current.
Pulse width
100 ms ±10%
maximum rising and falling time
1 ms
Alarm Output
Alarm delay time from the monitor to
remote equipment
The alarm delay time from the monitor to remote equipment is ≤2 seconds,
measured at the monitor signal output connector.
Alarm signal sound pressure level range
45 db(A) to 85 db(A) within a range of one meter
Trends
•
Standard-capacity internal card: up to 120 hours of trend data with the
resolution no less than 1 minute, or up to 1200 hours of trend data with
the resolution no less than 10 minutes.
•
High-capacity internal card: up to 240 hours of trend data with the
resolution no less than 1 second, or up to 2400 hours’ trend data with the
resolution no less than 10 minutes.
Events
•
Standard-capacity internal card: 1000 events, including parameter
alarms, arrhythmia events, technical alarms, and so on.
•
High-capacity internal card: 2000 events, including parameter alarms,
arrhythmia events, technical alarms, and so on.
NIBP measurements
•
Standard-capacity internal card: 1000 sets.
•
High-capacity internal card: 3000 sets.
Interpretation of resting 12-lead ECG
results
20 sets
Full-disclosure waveforms
•
Standard-capacity internal card: up to 48 hours for one waveform. The
specific storage time depends on the waveforms stored and the number
of stored waveforms.
•
High-capacity internal card: up to 48 hours for all parameter waveforms.
ST view
A maximum of 120 hours of ST segment waveforms. One group of ST
segment waveforms is stored every minute.
OxyCRG view
Up to 48 hours of OxyCRG events.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...