D - 8
D.2.10
AG Technical Alarm Messages
CO2 Low Barometric
Low
C
The ambient pressure exceeds the operating
pressure range or CO
2
module fails.
1. Make sure that the ambient pressure meets the
specifications, and check for sources that affect
the ambient pressure.
2. Replug the module.
3. If the alarm persists, contact your service
personnel.
CO2 Airway Occluded
Low
C
1. Check if the sample line is kinked or occluded.
2. Replace the sample line.
3. Replug the module.
4. If the alarm persists, contact your service
personnel.
CO2 No Filterline
Low
A
Make sure that the filterline is connected.
CO2 Calibration Required
Low
C
Perform a calibration.
CO2 Airway Error
Low
C
1. Check if the sample line is kinked or occluded.
2. Replace the sample line.
3. Replug the module.
4. If the alarm persists, contact your service
personnel.
CO2 Adapter Error
Low
A
Check, clean or replace the airway adapter.
Perform a zero calibration.
CO2 No Sensor
Low
A
Make sure that the CO
2
transducer is connected.
CO2: Change Watertrap
Low
C
Replace the watertrap.
CO2 Watertrap Mismatch
Low
C
Check the patient category and use a correct
watertrap.
CO2: Change O2 Cell
Low
C
The oxygen sesnor is depleted or fails. Replace
the oxygen sensor.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
AG No Watertrap
Low
B
Check the connections of the watertrap and re-
connect it.
AG: Change Watertrap
Low
C
Replace the watertrap.
AG Watertrap and Patient
Mismatch
Low
C
Check the patient category and use a correct
watertrap.
AG Zero Failed
Low
C
There is external electromagnetic interference,
airway occlusion or module failure.
1. Check for external inference sources.
2. Check for “AG Airway Occluded” alarm
message. Remove the occlusion.
3. If the alarm persists, contact your service
personnel.
Anesthetic Mixture
Low
C
Anesthetic mixture is detected.
AG Airway Occluded
Low
C
1. Check if the sample line is occluded.
2. Check the sample line.
3. Replug the module.
4. If the alarm persists, contact your service
personnel.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...