Nurse Call
The monitor provides a nurse call connector to output nurse call signal when a user-defined alarm occurs. To
obtain nurse call signal, use the nurse call cable to connect the hospital nurse call system with the monitor’s
nurse call connector.
Alarms are indicated on the nurse call device only when the following conditions are met:
■
The nurse call system is enabled.
■
A user-defined alarm occurs.
■
Alarms are not paused or reset.
WARNING
•
Do not rely exclusively on the nurse call system for alarm notification. Remember that the most
reliable alarm notification combines audible and visual alarm indications with the patient’s clinical
condition.
9.11
Calling for Help
In case of needing a help, you can call monitors in the same department, the central station, and the nurse call
system from your monitor so that nearby doctors and nurses can come for help.
To call help, select the
Call Help
quick key and select
OK
from the popup dialog box. If you did not select
OK
, the
monitor will automatically send out the call help signal in five seconds.
After the call help signal is sent out, the
Call Help
quick key flashes in red. If you need to stop calling for help,
select the
Call Help
quick key again.
Monitors receiving the call help signal issue a sound and a dialog box pops up indicating which monitor is
calling. Select
OK
to acknowledge the call and stop the sound at this monitor.
NOTE
•
The call help function works only when the monitor is connected to the network.
•
The call help sound may disturb patients in the same department.
9.12
CPB Mode
The CPB (Cardiopulmonary Bypass) mode is activated only if you set the department to
OR
.
In the CPB mode, except for BIS related alarms, all the physiological alarms and technical alarms are switched off.
So when performing CPB, you can put the monitor in the CPB mode in order to inactivate unnecessary alarms.
9.12.1
Entering the CPB Mode
To enter the CPB mode, choose either of the following ways:
■
Select the
CPB Mode
quick key.
■
Select the
Main Menu
quick key
→
from the
Alarm
column select
CPB Mode
.
In the CPB mode,
CPB Mode
is displayed in the physiological alarm area with a red background color.
NOTE
•
When the CPB mode is entered, the monitor stops all NIBP measurements. You can restart NIBP
measurements after entering the CPB mode.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...