D - 6
D.2.6
NIBP Technical Alarm Messages
D.2.7
IBP Technical Alarm Messages
Note: XX represents an IBP label, for example PA, CVP, FAP, P1, and so on.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
NIBP Loose Cuff
Low
A
There is a leak in the cuff or air tubing. Use a cuff
of correct type based on the patient size. Apply
the cuff and connect the air tubing as instructed
in the manual.
NIBP Air Leak
Low
A
Check the NIBP cuff and pump for leakages.
NIBP Airway Error
Low
A
The air tubing may be occluded. Check the air
tubing for an occlusion or kinking. If the alarm
persists, contact your service personnel.
NIBP Weak Signal
Low
A
The patient’s pulse is weak or the cuff is loose.
Check the patient’s condition and replace the cuff
application site.
NIBP Overrange
Low
A
The measured NIBP value exceeds the module
measurement range. Check the patient’s
condition.
NIBP Excessive Motion
Low
A
Check the patient’s condition and reduce patient
motion.
NIBP Cuff Overpress
Low
A
The NIBP airway may be occluded. Check the
airway and measure again. If the alarm persists,
contact your service personnel.
NIBP Timeout
Low
A
The measurement time exceeds 120 seconds in
the adult or pediatric mode, or exceeds 90
seconds in the neonatal mode, and the BP value
cannot be obtained. Check the patient’s
condition and NIBP connections, or replace the
cuff and measure again.
NIBP Cuff and Patient
Mismatch
Low
A
The cuff type mismatches the patient category.
Verify the patient category or replace the cuff if
necessary. If patient catergory is correct, check
that the tubing is not bent and the airway is not
occluded.
NIBP Airway Leak
Low
A
Airway leakage is found during the NIBP leakage
test. Check the NIBP cuff and pump for leakages.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
XX Sensor Error
Med
C
The IBP sensor fails. Replace the sensor.
XX No Sensor
High, Med, or
Low,
configurable
A
The IBP patient cable and/or corresponding IBP
sensor is not connected or detached. Check the
cable and sensor connection.
XX No Pulse
Low
A
The catheter may be occluded. Please flush the
catheter.
XX Disconnected
High
C
The liquid way is disconnected from the patient,
or the three-way valve is open to the air. Check
the connection of the liquid way, or check the
valve is open to the patient. If the alarm persists,
contact your service personnel.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...