13 - 7
To set the averaging time, follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
SpO2 Setup
tab.
3.
Set
Averaging
.
13.6.8
Changing Sensitivity (for Mindray SpO
2
)
The SpO
2
value displayed on the monitor screen is the average of data collected within a specific time. The
shorter the averaging time is, the quicker the monitor responds to changes in the patient’s oxygen saturation
level. Contrarily, the longer the averaging time is, the slower the monitor responds to changes in the patient’s
oxygen saturation level, but the SpO
2
measurement is more stable. For critically ill patients, selecting shorter
averaging time will help understanding the patient’s state.
To set the averaging time, follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
SpO2 Setup
tab.
3.
Select
Sensitivity
, and then toggle between
High
,
Med
and
Low
, which respectively correspond to 7 s, 9 s
and 11 s.
13.6.9
Showing/Hiding PI
You can set whether to display PI in the SpO
2
parameter area. To do so, follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
1.
Select the
Setup
tab.
2.
Switch on or off
Display PI.
13.6.10 Monitoring SpO
2
and NIBP Simultaneously
When monitoring SpO
2
and NIBP on the same limb simultaneously, you can switch on
NIBP Simul
to lock the
SpO
2
alarm status until the NIBP measurement ends. If you switch off
NIBP Simul
, low perfusion caused by NIBP
measurement may lead to inaccurate SpO
2
readings and therefore cause false physiological alarms.
To set the
NIBP Simul
, follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
Alarm
tab.
3.
Select
SpO2
tab.
4.
Set
NIBP Simul
.
13.6.11 Changing the Sweep Speed of the Pleth Wave
To set the sweep speed of Pleth waveform, follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
SpO2 Setup
tab.
3.
Set
Speed
.
13.7
Changing the PR Settings
13.7.1
Changing the PR Alarm Settings
To change the PR alarm settings, follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
PR Alarm
tab.
3.
Enter the password if required.
Содержание ePM 10M
Страница 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 58: ...4 8 This page intentionally left blank ...
Страница 62: ...5 4 This page intentionally left blank ...
Страница 118: ...11 4 This page intentionally left blank ...
Страница 134: ...13 12 This page intentionally left blank ...
Страница 144: ...15 8 This page intentionally left blank ...
Страница 156: ...16 12 This page intentionally left blank ...
Страница 174: ...18 12 This page intentionally left blank ...
Страница 182: ...19 8 This page intentionally left blank ...
Страница 192: ...20 10 This page intentionally left blank ...
Страница 222: ...24 4 This page intentionally left blank ...
Страница 228: ...25 6 This page intentionally left blank ...
Страница 256: ...28 6 This page intentionally left blank ...
Страница 264: ...29 8 This page intentionally left blank ...
Страница 268: ...30 4 This page intentionally left blank ...
Страница 280: ...31 12 This page intentionally left blank ...
Страница 346: ...E 4 This page intentionally left blank ...
Страница 350: ...F 4 This page intentionally left blank ...
Страница 360: ...G 10 This page intentionally left blank ...
Страница 361: ...H 1 H Declaration of Conformity ...
Страница 362: ...H 2 This page intentionally left blank ...
Страница 363: ......
Страница 364: ...P N 046 012607 00 6 0 ...