4-2 Product Principle
The management functions of the probe decouple the transmitting/receiving channel, and
communicate with the engine board.
As the independent sub-module, 4D driving function and physiological signals are
subordinate to the engine board.
As an option, CW receiving function can be realized through an independent single board
and is subordinate to TRA board signal.
Back-end unit: more functions are achieved on back-end unit. It basically resembles with PC. It
is equipped with main functions of the main unit, calculation capacity of image post-process,
operator-system interface to human-computer interaction, etc. PC module takes charge of
back-end unit. The elements are shown above.
Power supply unit: it supplies the power to the front-end unit, back-end unit and peripherals.
Other elements are shown above.
PHV module is subordinate to DC-DC module.
DC-DC module includes DC-DC board and battery assembly.
AC module includes AC-DC board, AC interface, control adapter board, etc.
4.2 Ultrasound Front-end Unit
Figure 4-2 Schematic diagram of ultrasound system front-end
Front-end unit mainly consists of:
Engine board
TR board
Probe board
Содержание DC-68
Страница 2: ......
Страница 12: ......
Страница 18: ......
Страница 30: ......
Страница 45: ...System Installation 3 15...
Страница 54: ......
Страница 72: ......
Страница 90: ......
Страница 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Страница 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Страница 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Страница 134: ......
Страница 180: ......
Страница 212: ......
Страница 254: ......
Страница 258: ......
Страница 280: ......
Страница 281: ...P N 046 011984 00 7 0...