A-16 Electrical Safety Inspection
ELECTRICAL SAFETY INSPECTION
8-
Patient Auxiliary Current
NOTE
If the current test standard being used does not include Patient Auxiliary Current DC readings,
or the DC option is not enabled, then DC readings will not be available through the APPLIED PART
SOFT KEY selections.
Failure
Check any broken of the AC cable. Replace a new one if any portion defective.
Check any broken of the enclosure. Replace any defective part.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user
or owner to correct any deviations. As a work around, check the other outlets to see if they could be
used instead.
Change another probe to confirm if the fail is caused by console.
Inspect wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if situation can not be
corrected, submit a Safety Failure Report to document the system problem. Remove unit from
operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and disposal.
LIMITS
All countries
For BF ECG input and transducer
100μA
Normal Condition
500μA
Single Fault Condition
Содержание DC-68
Страница 2: ......
Страница 12: ......
Страница 18: ......
Страница 30: ......
Страница 45: ...System Installation 3 15...
Страница 54: ......
Страница 72: ......
Страница 90: ......
Страница 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Страница 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Страница 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Страница 134: ......
Страница 180: ......
Страница 212: ......
Страница 254: ......
Страница 258: ......
Страница 280: ......
Страница 281: ...P N 046 011984 00 7 0...