C-10 Description of Self Test Items
Suggestion to failure test
Restart the device to perform the self test if the test result appears
Error.
It is necessary to restore
the device (OS+doppler) if
Error
re-appears.
It is recommended to replace control panel if the test result is
FAIL
.
C.1.25
Z0802
Control Panel Key, Encoder, TGC, Trackball
Function Test
Top test item
USB communication test of control panel
Test content
Test whether the control panel keys, encoders, TGCs, trackball, backlight LEDs and indicating
LEDs on the keyboard work well, and read board ID, keyboard FPGA version and program
version.
Test description
The keyboard test interface is shown as follows:
As shown in the figure above, click [Start] to perform the keyboard test. The controls on the
simulation keyboard correspond with those on the real keyboard assembly.