13-2 Troubleshooting of Regular Malfunctions
13.1.3 Troubleshooting as the System Unable to Power On
No.
Fault Description
Cause Analysis
Solution
1
AC power indicator remains
off;
1 No AC input;
2 Errors on AC-DC assembly
1 Check the connection
of AC
2 Replace AC-DC
assembly
2
AC indicator remains: on;
Power on/off indicator: the
indicator is off or does not
blink after pressing the button.
Power status indicator: 3.3
V/5 V/12 V is off;
The possibility of the failure on
PC carrier board reaches 90%.
Replace PC board
3
AC indicator remains: on;
Power on indicator: green
indicator blinks after pressing
the power button;
Power status indicator: 3.3
V/5 V/12 V is off;
PC carrier board responds with
the power on/power off
information. PC module does not
respond with power on/power off
information.
The possibility of the failure on
PC module reaches 90%.
Replace PC module
4
AC indicator remains: on;
Power on indicator: yellow
indicator starts to blink after
pressing the power button.
DC-DC board fails to respond.
The possibility of the failure on
DC-Dc board reaches 90%.
Replace DC-DC board
Содержание DC-68
Страница 2: ......
Страница 12: ......
Страница 18: ......
Страница 30: ......
Страница 45: ...System Installation 3 15...
Страница 54: ......
Страница 72: ......
Страница 90: ......
Страница 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Страница 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Страница 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Страница 134: ......
Страница 180: ......
Страница 212: ......
Страница 254: ......
Страница 258: ......
Страница 280: ......
Страница 281: ...P N 046 011984 00 7 0...