System Diagnosis and Support 11-17
6. Self test status display
The version and release date of the self test software appear on the top left of the status bar when
the self test program is running. DC-70 SelfTest Software for Maintenance Version: xx; Release
Date: YYYY-MM-DD.The software version in the status bar becomes the name of current test
during the test.The current test progress and overall test progress are displayed simultaneously on
the status bar.
Beyond that, the status bar shows the remaining time of the test.
7. Monitoring information
Monitoring information include: current time, CPU occupation rate, CPU 1 temperature, CPU 2
temperature, CPU 3 temperature, CPU 4 temperature, PCT1 temperature, PCT2 temperature and
PCT3 temperature.
If the temperature exceeds the working temperature, the temperature status in progress bar
appears in red.
8. System dialog box
The color of the element in the system dialog box changes as the color of test result changes in the
test process. If the test fails, the element becomes red
FAIL
; if the test succeeds, it becomes green
PASS
. If the color of the element keeps the same, it indicates the test is not proceeded.
9. Test information
The test result is displayed after the test is completed. Green
PASS
indicates the pass of the
sub-item. Red
FAIL
indicates the failure of the sub-item. If the item has not been tested, it appears
in grey
Skip.
If the board and the assembly are not in the place, it appears in grey
NaN
. Red
Error
appears as the test item error occurs. The test result of parent test item is the aggregate for the test
results of all selected sub-test items. If there is
FAIL
test in the selected list, the test result of parent
test item is
FAIL
. As long as the test result of the selected item is
PASS
, the test result of parent test
item would be
PASS
.As shown in the following figure.
Содержание DC-68
Страница 2: ......
Страница 12: ......
Страница 18: ......
Страница 30: ......
Страница 45: ...System Installation 3 15...
Страница 54: ......
Страница 72: ......
Страница 90: ......
Страница 108: ...8 2 Field Replaceable Unit 8 1 Explosive View A0 B0 C0 D0 E0 F0 G0...
Страница 121: ...Field Replaceable Unit 8 15 8 2 5 Main Unit Assembly E0 E2 E4 E6 E5 E9 E8 E3 E7 E1...
Страница 123: ...Field Replaceable Unit 8 17 E13 E22 E10 E17 E16 E14 E18 E19 E23 E20 E21 E15 E11 E12 E24...
Страница 134: ......
Страница 180: ......
Страница 212: ......
Страница 254: ......
Страница 258: ......
Страница 280: ......
Страница 281: ...P N 046 011984 00 7 0...