Basic Operations
3-3
3.1.2 Environmental Requirements
The operating environment of the patient monitor must meet the requirements specified in
this manual.
The environment where the patient monitor is used shall be reasonably free from noises,
vibration, dust, corrosive, flammable and explosive substances. If the patient monitor is
installed in a cabinet, sufficient space in front and behind shall be left for convenient
operation, maintenance and repair. Moreover, to maintain good ventilation, the patient
monitor shall be at least 2 inches (5cm) away from around the cabinet.
When the patient monitor is moved from one place to another, condensation may occur as a
result of temperature or humidity difference. In this case, never start the system before the
condensation disappears.
WARNING
z
Make sure that the operating environment of the patient monitor meets the specific
requirements. Otherwise unexpected consequences, e.g. damage to the equipment,
could result.
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...