Product Specifications
A-8
Output mode
Relay
Electrical requirements
≤
60W,
≤
2A,
≤
36VDC,
≤
25VAC
Isolation voltage
1500 VAC
Contact type
Normally open or normally contact (optional)
Defib Sync Pulse
Output impedance
50
Ω
Max time delay
35 ms (R-wave peak to leading edge of pulse)
Amplitude
High level: 3.5 to 5 V, providing a maximum of 1 mA output
current;
Low level: < 0.5 V, receiving a maximum of 5 mA input current.
Pulse width
100 ms ±10%
Limited current
15 mA rating
Rising and falling time
<1 ms
Digital video output (DVI-D connector)
Video signals
Single Link TMDS
DDC
Signals 12C compliant
A.4 Data Storage
Trends
Trends: 120 hours, at 1 min resolution
Minitrends: 1 hour, at 1 s resolution
Parameter alarms
100 alarms and related parameter waveforms. The waveform
recording length can be 8s, 16s or 32s.
Arrh. events
100 arrhythmia events and relate waveforms. The waveform
recording length can be 8s, 16s or 32s.
NIBP measurements
1000 sets
12-lead analysis results
40 sets
Full-disclosure waveforms
The storage time depends on the waveforms stored and the
number of stored waveforms.
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...