Monitoring SpO
2
8-7
8.5.6 Setting the Pulse Rate
In the [
SpO2 Setup
] menu, select [
PR Setup
] to enter the [
PR Setup
] menu. In the [
PR
Setup
] menu, you can:
Select [
PR
] and then select either:
[
On
] to display the PR parameter window, or
[
Off
] to hide the PR parameter window.
Select [
Alm Source
] and then select either:
[
HR
]: if you want the HR to be the alarm source.
[
PR
] if you want the PR to be the alarm source.
[
Auto
]: If the [
Alm Source
] is set to [
Auto
], the monitor will use the heart rate
from the ECG measurement as the alarm source whenever a valid heart rate is
available. If the ECG signal is too weak, the monitor will automatically switch to
PR as the alarm source. As soon as the ECG signal recovers, the monitor will
automatically switch back to the heart rate as the alarm source.
8.5.7 Changing the Speed of the Pleth Wave
In the [
Pleth Waveform
] menu, select [
Sweep
] and and then select the appropriate setting.
The faster the waveform sweeps, the wider the waveform is.
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...