Monitoring IBP
11-7
Follow this procedure to perform the pressure calibration:
1. Disconnect the pressure transducer from the patient. Connect the 3-way stopcock, the
sphygmomanometer and the balloon pump through a T-shape connector, as shown
below.
2. Zero the transducer. After a successful zero, open the stopcock to the manometer.
3. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
Cal
.
IBP Press. >>
]. In the [
Cal. IBP
Press.
] menu, enter calibration values for calibrated pressures.
4. Inflate using the balloon pump until the mercury in the manometer reaches to the preset
calibration pressure value.
5. Adjust the preset calibration value until it equals to the reading on the manometer.
6. Select
the
[
Calibrate
] button to the right of the calibrated pressure. The monitor starts a
calibration.
7. When the calibration is completed, the message [
Calibration Completed!
] is displayed.
If the calibration failed, a prompt message will be displayed.
8. After the calibration is completed, disconnect the blood pressure tubing and the T-shape
connector.
Manometer
T-shape connector
3-way stopcock
Pressure transducer
Pressure adapter cable
IBP module
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...