15-1
15
Monitoring ICG
15.1 Introduction
Impedance cardiography (ICG) measures a patient’s hemodynamic status using a safe,
non-invasive method based on thoracic electrical bioimpedance (TEB) technology. ICG uses
four pairs of sensors to transmit a small electrical signal through the thorax. As velocity and
volume of blood in the aorta change, the ICG measures the changes in impedance from
systole to diastole to calculate hemodynamic parameters.
15.2 Safety
WARNING
z
Apply ICG monitoring to adult patients in height of 122 to 229 cm, weight of 30 to
159 kg only.
z
ICG monitoring should not be used concurrently on patients with minute
ventilation pacemakers when the MV sensor function is activated.
z
During ICG monitoring, make sure that the conductive paste on the ICG sensors
never come into contact with other conductive parts.
z
ICG sensors are for single patient use only.
Connector for ICG cable
Setup key
Indicator
Check sensor
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...