Alarms
6-9
numeric and violated alarm limit stop flashing as soon as the initial alarm condition goes
away.
To set alarms to latching or non-latching:
1. Select
the
[
Alarm Setup
] QuickKey, or [
Main Menu
]
→
[
Alarm Setup >>
].
2. Select
[
Latching Alarms
] and toggle between [
Yes
] and [
No
].
Technical alarms are always non-latching. You can clear the latched alarms by pressing the
hardkey.
6.9 Clearing Technical Alarms
For some technical alarms, their alarm lamp flashing and alarm tones are cleared and the
alarm messages change to prompt messages after the
hardkey is pressed. After the patient
monitor restores the normal status, the patient monitor can give alarm indications correctly
when these alarms are triggered again.
For some other technical alarms, all their alarm indications are cleared after the
hardkey
is pressed. After the patient monitor restores the normal status, the patient monitor can give
alarm indications correctly when these alarms are triggered again.
6.10 When an Alarm Occurs
When an alarm occurs, observe the following steps and take proper actions:
1. Check the patient’s condition.
2. Confirm the alarming parameter or alarm category.
3. Identify the source of the alarm.
4. Take proper action to eliminate the alarm condition.
5. Make sure the alarm condition is corrected.
For troubleshooting specific alarms, see appendix
Alarm Messages
.
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...