Product Specifications
A-14
A.5.4 NIBP
Standards
Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1,
EN1060-3 and SP10
Technique Oscillometric
Mode of operation
Manual, Auto and STAT
Auto mode repetition intervals 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240 or 480 min
STAT mode cycle time
5 min
Adult, pediatric:
180 s
Max measurement time
Neonate:
90 s
Heart rate range
40 to 240 bpm
Adult
Pediatric
Neonate
Systolic:
40 to 270
40 to 200
40 to 135
Diastolic:
10 to 210
10 to 150
10 to 100
Measurement ranges
(mmHg)
Mean:
20 to 230
20 to 165
20 to 110
Accuracy
Max mean error:
±
5 mmHg
Max standard deviation: 8 mmHg
Resolution 1mmHg
Initial cuff inflation pressure
Adult:
Pediatric:
Neonate:
178±5 mmHg
133±10 mmHg
87±5 mmHg
Software overpressure
protection
Adult:
Pediatric:
Neonate:
297±3 mmHg
240±3 mmHg
147±3 mmHg
Hardware overpressure
protection
Adult:
Pediatric:
Neonate:
330±5 mmHg
330±5 mmHg
165±5 mmHg
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...