Maintenance
25-6
If the difference is great, you should perform a calibration. Select [
Calibrate >>
] to
enter the calibrate menu.
6. Enter the vented gas concentration. If you use only one gas for calibration, set other
gases’ concentration to 0.
7. Select
[
Calibrate
] to start calibration.
8. If the calibration is finished successfully, the message [
Calibration Completed!
] is
displayed. If the calibration failed, the message [
Calibration Failed!
] is displayed.
Perform another calibration.
9. Select to exit the current menu.
25.8 Setting up IP Address
1. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance
>>]
→
enter the
required password.
2. Select
[
IP Address Setup >>
] to change IP address settings.
If the patient monitor is connected to a CMS, its IP address should be set up. The user should
not change the patient monitor’s IP address randomly. If you want to know details about IP
address setup, contact the technical personnel in charge of the CMS.
25.9 Entering/Exiting Demo Mode
To enter the Demo mode:
1. Select
[
Main Menu
]
→
[
Maintenance >>
].
2. Select
[
Demo >>
]. Enter the required password and then select [
Ok
].
To exit the Demo mode:
1. Select
[
Main Menu
]
→
[
Maintenance >>
].
2. Select
[
Exit Demo
] and then select [
Ok
].
3. The patient monitor exits the Demo mode and then shuts down automatically.
WARNING
z
The Demo mode is for demonstration purpose only. To avoid that the simulated
data are mistaken for the monitored patient’s data, you must not change into
Demo mode during monitoring. Otherwise, improper patient monitoring and
delayed treatment could result.
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...