Monitoring SpO
2
8-4
8.5 Changing SpO
2
Settings
8.5.1 Accessing SpO
2
Menus
By selecting the SpO
2
parameter window, you can access the [
SpO2 Setup
] menu.
By selecting the PR numeric, you can access the [
PR Setup
] menu.
By selecting the Pleth wave, you can access the [
Pleth Waveform
] menu.
8.5.2 Adjusting the Desat Alarm Limit
The desat alarm is a high level alarm notifying you of potentially life threatening drops in
oxygen saturation. To adjust the desat alarm limit, select [
Desat Limit]
in the [
SpO2 Setup
]
menu and then adjust the limit. When the SpO
2
value is below the desat alarm limit, the
message [
SpO2 Desat]
is displayed. In this case, the monitor will not alarm if other
physiological alarms occur.
8.5.3 Setting SpO
2
Sensitivity
To set the SpO
2
measurement sensitivity for MPM SpO
2
module, select [
Sensitivity
] in the
[
SpO2 Setup]
menu and toggle between [
High
], [
Med
] and [
Low
].
[
High
]: The monitor gives quick response to changes in the oxygen saturation level, but
the measurement accuracy might be compromised. When monitoring critically ill
patients, it is recommended that the measurement sensitivity is set to [
High
] so that you
can timely understand the patient’s state of illness.
[
Low
]: The monitor gives slow response to changes in the oxygen saturation level, but
the measurement accuracy is relatively high.
For Masimo SpO
2
modules, you can set the SpO
2
measurement sensitivity to [
Normal
] or
[
Maximum
]. The [
Normal
] and [
Maximum
] sensitivities respectively correspond to the
[
High
] and [
Low
] sensitivities mentioned above.
Содержание BeneView T8
Страница 1: ...BeneView T8 Patient Monitor Operation Manual...
Страница 2: ......
Страница 8: ...2 FOR YOUR NOTES...
Страница 24: ...Safety 1 6 FOR YOUR NOTES...
Страница 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Страница 74: ...Alarms 6 10 FOR YOUR NOTES...
Страница 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Страница 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Страница 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Страница 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Страница 184: ...Review 19 10 FOR YOUR NOTES...
Страница 206: ...Recording 21 8 FOR YOUR NOTES...
Страница 210: ...Other Functions 22 4 FOR YOUR NOTES...
Страница 256: ...Product Specifications A 24 FOR YOUR NOTES...
Страница 290: ...P N 6800 20 50657 1 0...