Page 30
Instruction Manual Software
thicknessCONTROL MTS 8202.LLT / 4350127.256
5.4
View System Overview
Directly beneath the menu the current measurement values, data and information of the
measuring system are displayed. In view
System overview
, selectable via view tab,
additional information of the measuring system is displayed.
The segment
Calibration target position
shows the current state of the
calibration target. In manual mode the calibration target can be swiveled in/out in the
calibration gap through the
Move to parking position
and
Move to calibra-
tion position
buttons.
In segment
Automatic axis motor
the
same kind of display applies to the limit
switches. Additionally, the horizontal bar view
describes the current position of the automatic
axis. The manual commands are activated and
deactivated via the menu
Measurement >
Manual mode
. The displays of the digital in-
puts show the current state of the servomotor.
Examples symbols, automatic mode
Measurement carriage not in
horizontal motor zero position.
Measurement carriage travers-
es in maximal position direc-
tion.
Measurement carriage in hori-
zontal parking position.
The segment
Digital inputs
shows the
state of the measuring system by means of
virtual LEDs. E.g. area stop, main switch, USV,
compressed air, air preparation filter, safety
mechanisms ASO are surveilled. Additionally,
the current speed and length
are displayed.
Example symbols
Process OK
Process ordinary
Process critical
In the array
Status
the current system errors are displayed in words. These system
errors are saved in the window of the system logger.
Fig. 33 System overview
Содержание 4350127.256
Страница 1: ...thicknessCONTROL MTS 8202 LLT C frame Non contact thickness measurement Operating Instructions...
Страница 10: ...Page 10 Overview Technical Data Safety Instructions thicknessCONTROL MTS 8202 LLT 4350127 256...
Страница 26: ...Page 26 Installation and Commissioning Instructions thicknessCONTROL MTS 8202 LLT 4350127 256...
Страница 56: ...Page 56 Declaration of Incorporation thicknessCONTROL MTS 8202 LLT 4350127 256...
Страница 78: ...Operating Instructions scanCONTROL 29xx...
Страница 82: ...scanCONTROL 29xx Appendix A 1 Accessories 62 A 1 1 Recommended Accessories 62 A 1 2 Optional Accessories 63...
Страница 83: ......
Страница 119: ...Page 42 Operation of the Sensor with a PC scanCONTROL 29xx Fig 23 Predefined measuring fields...