Measuring Field Selection
The optical design of the sensor satisfies the so-called “Scheimpflug condition” which ensures optimum
mapping over the complete measuring range. In doing so, the measuring range is mapped on a rectangu-
lar matrix. The distortions resulting from this are shown, see
. The usable measuring range is always
trapezoidal.
The assigned maximum x-values for the z-coordinates can be found, see
.
Please refer to the sensor acceptance report of your sensor. A slight range shifting of a measuring field is
possible and depends on the sensor.
The top edge corresponds to the start of the measuring range and the bottom edge to the end of the measur-
ing range. The corners of the predefined measuring fields are on a grid with grid spacing of 1/8 of the matrix.
The sensor matrix used in the scanCONTROL 29xx supports the reading of a restricted measuring field. The
following picture, see
, shows the predefined view areas and the associated measuring fields.
Содержание 4350127.256
Страница 1: ...thicknessCONTROL MTS 8202 LLT C frame Non contact thickness measurement Operating Instructions...
Страница 10: ...Page 10 Overview Technical Data Safety Instructions thicknessCONTROL MTS 8202 LLT 4350127 256...
Страница 26: ...Page 26 Installation and Commissioning Instructions thicknessCONTROL MTS 8202 LLT 4350127 256...
Страница 56: ...Page 56 Declaration of Incorporation thicknessCONTROL MTS 8202 LLT 4350127 256...
Страница 78: ...Operating Instructions scanCONTROL 29xx...
Страница 82: ...scanCONTROL 29xx Appendix A 1 Accessories 62 A 1 1 Recommended Accessories 62 A 1 2 Optional Accessories 63...
Страница 83: ......
Страница 119: ...Page 42 Operation of the Sensor with a PC scanCONTROL 29xx Fig 23 Predefined measuring fields...