Page 22
Installation and Commissioning Instructions
thicknessCONTROL MTS 8202.LLT / 4350127.256
4.4
Overview Interfaces
1 .. 6
7
8
Fig. 23 Interface connections required from the system operator
1
120V / 60Hz / 10A
7
Earthing
2
Emergency stop line
8
customer air
3
Remote support (VPN)
4
Interface: Ethernet IP
5
4x Analog output
6
Earthing
not included in delivery
4.5
Electrical Power Supply
The measuring system requires an AC mains voltage of 120 VAC, ±5 % / 60 Hz for proper operation.
Connect the measuring system to the supply voltage.
The measuring system must be connected as follows:
Connect the cable for the 120V / 60Hz power to +ET1-1X0:
Signal
-1X0
contact
Customer
cable
L1 (120V AC)
1
N
2
PE
3
Connect the Emergency Stop connection to -7X1:
Signal
-7X1
contact
Customer
cable
emergency Stopp
contact : 1
1
emergency Stopp
contact : 1
2
emergency Stopp
contact : 2
3
Содержание 4350127.256
Страница 1: ...thicknessCONTROL MTS 8202 LLT C frame Non contact thickness measurement Operating Instructions...
Страница 10: ...Page 10 Overview Technical Data Safety Instructions thicknessCONTROL MTS 8202 LLT 4350127 256...
Страница 26: ...Page 26 Installation and Commissioning Instructions thicknessCONTROL MTS 8202 LLT 4350127 256...
Страница 56: ...Page 56 Declaration of Incorporation thicknessCONTROL MTS 8202 LLT 4350127 256...
Страница 78: ...Operating Instructions scanCONTROL 29xx...
Страница 82: ...scanCONTROL 29xx Appendix A 1 Accessories 62 A 1 1 Recommended Accessories 62 A 1 2 Optional Accessories 63...
Страница 83: ......
Страница 119: ...Page 42 Operation of the Sensor with a PC scanCONTROL 29xx Fig 23 Predefined measuring fields...