Functional Description
21
b) from the screen
Click at top right in the screen, on the Time Base zone (see opposite).
Description below of the Y(t) - Y(f) - XY display modes
1. Y(t): temporal view
of a waveform
Settings from 1ns to 200s
No averaging
Averaging coeff.
2
Averaging coeff.
4
Averaging coeff.
16
Averaging coeff.
64
Selection of a coefficient in order to calculate an average on the
samples displayed: this can be used for example to attenuate the
random noise observed in a signal.
For the averaging coefficient to be taken into account in the
representation of the signal, the "Repetitive signal" option must be
selected.
The calculation is done using the following formula:
Pixel N = Sample*1/Averaging rate + Pixel N-1
(1-1/Averaging rate):
Sample
Value of the new sample acquired at abscissa t
Pixel N
Ordinate of the pixel at abscissa t on the screen,
at instant N
Pixel N-1
Ordinate of the pixel at abscissa t on the screen,
at instant N-1
Vector
A vector is plotted between samples.
Envelope
The minimum and maximum observed at each horizontal position on
the screen are displayed. Use this mode to display a variation in time or
of amplitude, or a modulation.
The entire
acquisition
The whole of the acquisition (100,000 samples) is displayed on the
screen and a vector is plotted between
samples. Use this mode to
display all details of the acquisition. This function can be used on a
memory or on a curve already acquired.
Increased time resolution of a trace for a periodic signal.
If this option is checked, the signal can be averaged.
For time bases finer than 100µs/div. (without active zoom mode), the signal displayed
is reconstituted from several acquisitions. The time resolution can be as fine as 40ps.
If the signal is not repetitive, do use not this option. The time resolution will then be
±1ns.
If this choice is checked, reconstruction of the signal can take a rather long time.
The following parameters influence this time:
the time base
the frequency of recurrence of the trigger
the activity of the Averaging mode
During this reconstruction, the signal must be stable (amplitude, frequency, waveform).
To speed up the reconstruction following a change in the signal, stop the acquisition, then
restart: Stop/Run.