Technical characteristics
68
8. TECHNICAL CHARACTERISTICS
8.1. "Oscilloscope" function
Vertical deflection
Characteristics
OX 9062
OX 9102
OX 9104
OX 9304
Number of
channels
1
2
OX 9xx2:
2,
OX 9xx4:
4
Vertical
ranges
2.5mV to 200V/div.
Variation in steps (no continuously variable coefficient)
BW
to 3dB down
60MHz
100MHz
300MHz
Measured into a 50
Ω load with a signal having an amplitude of 6 div.
Max. input
voltage
2
1400 VDC, 1kVrms with the
Probix
HX0030 probe
Types of
input
Probix
safety connector: class 2, isolated inputs
Dynamic
of the vertical offset
±10 divisions in all ranges
Input
coupling
AC
DC
GND
10Hz to 60MHz
0 to 60MHz
reference
10Hz to 100MHz
0 to 100MHz
reference
10Hz to 300MHz
0 to 300MHz
reference
Bandwidth
limiters
at
≈
15MHz, 1.5MHz, 5kHz
Rise
time
in all vertical ranges. 2.5mV to
200V/div.
≈
5.85ns
≈
3.5ns
≈
1.17ns
Cross-talk
between channels
>70dB (
Same sensitivity in both channels)
Response
to rectangular
signals at 1kHz and 1MHz
Positive or negative overshoot
Overshoot
≤
4%
Vertical
resolution
of the display
±0.4% of full scale (without ZOOM)
0.025% in ZOOM mode (12 bits)
Accuracy
of the peak-to-peak gains
±2% with averaging from 4 to 1kHz
Accuracy
of the vertical measurements
in DC with offset
and averaging over 16
±[2.2% (reading) + 11% (sensitivity) + 40
0 μV]
applies to the following measurements: Vmin, Vmax, Vlow, Vhigh, Vavg, curs(1), curs(2)
Accuracy
of the vertical measurements
in AC without offset at 1kHz
with averaging over 16
±[2% (reading) + 1% (sensitivity)]
applies to the following measurements: Vamp, Veff, Dep+, Dep-
Resolution
of the measurements
12 bits
Accuracy
of the vertical offset
±[0,2% (reading) + 10% (sensitivity) + 40
0 μV]
Vertical
ZOOM
function on an acquired or
saved curve
ZOOM factors: 16 max.
Input
impedance
1 M
Ω
±0.5% approx. 12 pF
1
Instruments with two channels: CH1 and CH4, instruments with four channels: CH1, CH2, CH3, CH4
2
Refer to the figure (§ 9.4.3.): max. input voltage as a function of frequency
Only the assigned tolerance or limit values
are guaranteed values (after a half-hour
warm-up period). The values without
tolerances are given as an indication