Remote Programing
88
Special values
MAXimum
,
MINimum
are used to obtain the extreme values of the parameter.
UP
,
DOWN
are used to go to the value following or preceding the current state of the parameter.
Character strings
These are series of letters and/or digits enclosed in quotation marks " ".
Terminator
<NL>
We use
<NL>
as the general term designating a terminator.
NL
is the CR character (ASCII code 13 or 0x0D).
A command line must not be more than 80 characters long; it is terminated by a terminator.
Syntax of replies
The reply can comprise several elements separated by a comma ",". The last element is followed
by the terminator <NL>.
The data are of several kinds:
Key words
These are the same as those used as parameters, but here only the abbreviated form is
returned.
Digital values
Three formats are possible: NR1, NR2, and NR3.
Character string
There is no difference with respect to the parameters. If the string contains a key word, it is
returned in abbreviated form.