Functional Description
33
4.1.9. AUTOMATIC measurements, from the screen
Opening of the "Automatic
measurements" Menu window of the
channel
Opening of the "Automatic
measurements" Menu window of the 4
channels
The measurements are made and refreshed on the selected reference trace. All
measurements that can be made on this trace are displayed. (- . - -) is displayed for
measurements that cannot be made.
The window is closed by pointing to
with the stylus.
All
20 measurements
selected will be displayed in the status zone at the bottom of the screen,
on a ground the colour of the channel:
vmin
minimum peak voltage
trise
rise time
vmax
maximum
peak voltage
tfall
fall time
vpp
peak-to-peak voltage
wplus
positive pulse width
(at 50% of Vamp)
vlow
stabilized low voltage
wlow
negative pulse width
(at 50% of Vamp)
vhigh
stabilized high voltage
period
period
vamp
amplitude
freq
frequency
vrms
RMS voltage determined in the
measurement interval
dcycle
duty cycle
vrms_c
RMS voltage determined on a
whole number of cycles
npulses
number of pulses
vavg
mean voltage
over_pos
positive overshoot
sum
summation of the instantaneous
values of the signal
over_neg
negative overshoot
Measurement
conditions
The measurements are made on the part of the trace displayed on screen between
cursors T1 and T2.
Any modification of the signal entails an update of the measurements.
They are refreshed as the acquisition proceeds.
The accuracy of the measurements is optimum when at least two complete periods of
the signal are displayed.