Technical characteristics
78
8.4 "HARMONIC ANALYSIS" function
Presentation of the harmonics in bargraph form
Crosshair with vertical axis graduated in %
Horizontal axis graduated in orders of harmonic
Display of 63 orders
The harmonic analysis function can be implemented on the 4 channels
Display of the measurements made:
RMS level of the signal
total harmonic distortion with respect to the RMS value of the fundamental (THD).
RMS level of the harmonic selected
ratio in % of the RMS value of the selected harmonic to the RMS value of the fundamental
frequency of the selected harmonic
phase of the selected harmonic/fundamental
Harmonic analysis
Frequency of the fundamental
of the signal analyzed
from 40 to 450Hz
Condition
Accuracy of the measurements
In the domain of reference: 18°C to 28°C, at 50Hz and 60Hz
Level of the Fundamental
±(2% + 10 D)
Level of the Harmonics
±(3% + 10 D), ratio ±2%
ratio >4%
Harmonic distortion (THD)
±4%
Phase
±5%
ratio >4%
Variations in the nominal range of use 0°C to 40°C, at 50Hz and 60Hz
Level of the Fundamental
±(5%/10°C)
Level of the Harmonics
±(5%/10°C), ratio ±(1%/10°C)
ratio >4%
Harmonic distortion (THD)
±(5%/10°C)
Phase
±(10°/10°C)
ratio >4%